Языки

Processing of speech therapist tools

Pre-sterilization cleaning, disinfection and sterilization.

Probes are resistant to:

pre-sterilization purification consisting of preliminary rinse with running drinking water SSU 2874 for 0.5 ± 0.1 min, soaking in a solution of ethyl alcohol 76% at a temperature of 20 ± 10 ºС for 15 ± 1 min, washing in an aqueous solution containing ethyl alcohol not less than 76% for 3 ± 1 min, rinsing with distilled water SSU 6709 for 0.5 ± 0.1 min, drying with hot air at a temperature of 80 ± 5 ° C until the complete loss of moisture;

– air disinfection with dry hot air at a temperature of 95 ± 5 ° C;

– steam sterilization at a temperature of 100 ± 5 ° C for 10 ± 1 min.